Atomic Absorption Spectrometer (AAS)
Project |
Key Feature |
|
Optical System |
Wavelength Range |
(190~900)nm |
Spectral Bandwidth |
(0.1, 0.2, 0.4, 1.0, 2.0) nm five-stage automatic switching |
|
Monochromator |
C-T grating monochromator |
|
Wavelength Accuracy |
±0.1nm |
|
Wavelength Repeatability |
≤0.05nm |
|
Grating Line Density |
1800 lines/mm |
|
Flame Wavelength |
250nm |
|
Resolution |
Better than 0.1 nm |
|
Baseline Stability |
≤0.003 A/30 min (dynamic) ≤0.002 A/30 min (static) |
|
|
Flame Analysis |
Cu Characteristic Concentration |
≤0.02 μg/mL/1% |
Detection Limit |
≤0.003 μg/mL |
|
Precision RSD |
≤0.45% |
|
Burner Head |
Compatible with 100 mm full titanium burner head or 50 mm stainless steel burner head |
|
Position Adjustment |
Height and angle adjustment for optimal positioning, flame/hydride replacement completed within one minute |
|
Nebulizer |
Standard high-efficiency glass nebulizer and full titanium metal nebulizer; full titanium metal nebulizer is suitable for analyzing samples containing corrosive HF |
|
Graphite Furnace Analysis |
Cd Characteristic Concentration |
0.5×10-12g |
Detection Limit |
10×10-12g |
|
Temperature Control Range |
Room temperature to 2700°C |
|
Precision RSD |
≤3.0% |
|
Temperature Program |
Up to 20-stage heating program, with three heating methods: step, slope, and hold. |
|
Heating Rate |
≥2500℃/s |
|
Power Heating Rate |
≥2000℃/s |
|
Temperature Control Accuracy |
≤ 1% |
|
Temperature Reproducibility |
≤ 0.5% |